Field Safety Notice – Ethicon Sutures

8th August 2023

8th August 2023

Dear Customer,

Ethicon Johnson& Johnson Surgical Technologies is undertaking  Voluntary Product Recall recalling listed batches PDS™ II (polydioxanone) Suture & PDS™ Plus Antibacterial (polydioxanone) Suture which are detailed in the table below. You can also read the Field Safety notice by clicking here.

This voluntary removal stems from Ethicon receiving several complaints for one lot of undyed PDS™ II (polydioxanone) Sutures regarding low tensile strength. Internal testing on returned product from this lot confirmed that some PDS™ II (polydioxanone) Sutures from this lot did not meet Ethicon’s tensile strength requirement. An investigation identified  additional  lots  across  the  PDS™  Plus  Antibacterial  (polydioxanone) and  PDS™  II  (polydioxanone) Suture families that have the potential to be affected by the same root cause as the lot that received the complaints.

The voluntary recall has affected the following batches of 12 NVS product codes detailed within the tables below:


NVS Code Manufacture Code Batch/Lot Expiry date
548984 MPZ496H QCMDCS 28.02.2025
375287 PDP9625H RJMEKR 31.07.2023
514469 PDP9626H RKMMTB 31.08.2023
SAMHHE 31.12.2023
373503 PDP9861H SBMESR 31.01.2024
210845 W9624T SDMPKU 31.03.2027
9131 W9631T RLMAZJ 30.09.2026
SCMHJL 28.02.2027
105673 W9733T PM6878 31.10.2024
QEMCUE 30.04.2025
QEMEUB 30.04.2025
QGMHMB 31.05.2025
QGMHST 31.05.2025
QGMJQB 31.05.2025
QHMLLR 30.06.2025
QKMJDB 31.08.2025
QKMJDQ 31.08.2025
QKMMLZ 31.08.2025
QLMCLJ 30.09.2025
RAMBLL 31.12.2025
RAMBRT 31.12.2025
RAMBZX 31.12.2025
RCMAJX 28.02.2026
RCMBPQ 28.02.2026
RCMBUB 28.02.2026
RCMCBT 28.02.2026
RDMKQD 31.03.2026
RHMKQD 30.06.2026
RHMKUA 30.06.2026
RJMAJQ 31.07.2026
RJMASJ 31.07.2026
SHMMEX 30.06.2027
SHMMHT 30.06.2027
SMMBPC 31.10.2027
SMMCRC 31.10.2027
600040 W9734T QDMBUS 31.03.2025
QDMCKS 31.03.2025
RAMMXL 31.12.2025
REMMRE 30.04.2026
REMMXZ 30.04.2026
RGMEUD 31.05.2026
SLMAQH 30.09.2027
836862 W9740T RDMHBX 31.03.2026
209971 W9867T QDMBUQ 31.03.2025
225460 W9872T RAMMXU 31.12.2025
657384 W9957T PGZ717 31.05.2024
RCMKML 28.02.2026
RDMDHK 31.03.2026
RKMCKK 31.08.2026
SCMCRD 28.02.2027
502327 MPZ9626H SCMBTJ 28.02.2027
130771 W9626T RCMKMU 28.02.2026
RGMEDH 31.05.2026
RGMEKR 31.05.2026
RGMEUA 31.05.2026
SDMAZQ 31.03.2027
217566 W9861T SCMBAL 28.02.2027

 

What to do next – Please examine your stock immediately and take the following steps.

I have the affected product

  • Quarantine the product.
  • If you have any of the affected stock, then please contact our Customer Services Team on 01782 771111 (Option 3) to arrange the return of the products.
  • This affected product will only be accepted until 31st August 2023, after this date we will not accept this product.

I do not have this product

  • If you don’t have any of the products listed above, no further action is required.

If you require any further information, please contact our Customer Service Department on 01782 771 111.

Kind regards,

Andrew Duxbury

(Responsible Person)